AirLife Adult Manual Resuscitator, 40" (1.0 m) Oxygen Reservoir Tubing, Adult Mask, CO2 Detector, REF 2K8005C2

US FDA — Medical Devices · 2024-01-10 · United StatesClass I
Issuing authority
US FDA — Medical Devices
Alert date
2024-01-10
Country of alert
United States
Alert level
Class I
Alert type
recall
Barcode (EAN/GTIN)
50190752114164
Brand on record
Vyaire Medical
Product category
Medical Device

What the authority reported

Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.