Official recall record
Allura Xper FD10F; Model Number: 722002;
- Issuing authority
- US FDA — Medical Devices
- Recall initiated by firm
- 2025-12-15
- Country of alert
- Netherlands
- Alert level
- Class II
- Alert type
- recall
- Brand on record
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
What the authority reported
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Record last checked against the registry on 2026-05-31.