Allura Xper FD10; Model Number: 722003;

US FDA — Medical Devices · 2025-12-15 · NetherlandsClass II
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2025-12-15
Country of alert
Netherlands
Alert level
Class II
Alert type
recall
Brand on record
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

What the authority reported

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.