ProductGuru Recalls

Official recall record

Accessory Sets Syringes: 6ml syringe with a female luer with 15cm sealed SCD tubing, REF: RF-T15, and 60ml syringe on a female luer with 15c

US FDA — Medical Devices · 2023-12-19 · United StatesClass II
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2023-12-19
Country of alert
United States
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
10816203020205
Brand on record
OriGen Biomedical, Inc.
Product category
Medical Device

What the authority reported

Accessory sets contain a syringes not yet qualified for the accessory's intended purpose; due to design changes to the syringe barrels and a polypropylene resin change that resulted in loss of gamma compatibility, for accessory set syringes that were gamma sterilized, which resulted in yellowed syringes.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: US FDA — Medical Devices · United States · all filters