ProductGuru Recalls

Official recall record

SenTiva, REF: 1000; SenTiva Duo, REF: 1000-D; Generators Part of the VNS Therapy System. Used for Vagus Nerve Stimulation (VNS).

US FDA — Medical Devices · 2024-12-12 · United StatesClass II
Issuing authority
US FDA — Medical Devices
Alert date
2024-12-12
Country of alert
United States
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
05425025750405
Brand on record
LivaNova USA, Inc.

What the authority reported

Vagus nerve stimulator generators may stop delivering stimulation due to an internal, mechanically activated component that may become stuck in a closed position, which may result in patients returning to baseline seizure frequency or depressive symptoms and in the longer term, patients may be required to undergo generator replacement surgery to replace their device.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.