Official recall record
MOSAIQ Oncology Information System with Particle Therapy License (License Kit 45700000000IQRO)
- Issuing authority
- US FDA — Medical Devices
- Recall initiated by firm
- 2025-12-10
- Country of alert
- United States
- Alert level
- Class II
- Alert type
- recall
- Barcode (EAN/GTIN)
- 07340201500026
- Brand on record
- Elekta, Inc.
What the authority reported
Using oncology information system that manages workflow may result in overtreatment if:Plan delivered with Particle Therapy IHE-RO TDW-II interface,With one field per fraction or last field partial treatment,Partial treatment occurs followed by treatment field interruptions with manually recorded partial treatment,Then treatment unit doesn't check meterset value so partial treatment delivered twice
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Record last checked against the registry on 2026-05-31.
Same barcode in other registries
The same EAN appears in 1 other official recall record(s) — a cross-border signal worth checking if you bought this product abroad.
- 2025-01-15 · US FDA — Medical Devices
MOSAIQ Oncology Information System
Other recalls from this brand
- 2025-12-12 · US FDA — Medical Devices
Brand Name: Leksell GammaPlan Product Name: Radionuclide RT Treatment planning system Software Version: 11,3