ProductGuru Recalls

Official recall record

Brand Name: RayStation Intended as a Radiation Therapy Treatment Planning System. Model/Catalog Number: 7.0.0.19, 8.0.0.61, 8.0.1.10 Sof

US FDA — Medical Devices · 2025-01-16 · SwedenClass II
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2025-01-16
Country of alert
Sweden
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
07350002010068
Brand on record
RAYSEARCH LABORATORIES AB

What the authority reported

Inconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT calibration curve used in RayStation version 4.5 to 2024B including some service packs.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: US FDA — Medical Devices · Sweden · all filters

Same barcode in other registries

The same EAN appears in 1 other official recall record(s) — a cross-border signal worth checking if you bought this product abroad.

Other recalls from this brand

All 15 official recall notices naming RAYSEARCH LABORATORIES AB on this site →

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.