VITROS XT 7600 Integrated System, Catalog No. 6844461

US FDA — Medical Devices · 2024-01-26 · United StatesClass II
Issuing authority
US FDA — Medical Devices
Alert date
2024-01-26
Country of alert
United States
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
10758750031610
Brand on record
Ortho-Clinical Diagnostics, Inc.
Product category
Medical Device

What the authority reported

A software defect affecting VITROS Systems running VITROS Software Versions 3.8.0 and 3.8.1 causes the Configurable Westgard Rules to not report as expected. When a new control is defined, configurable Westgard rules, if selected, are not flagged or reported with the applicable QC results if the statistical criteria is met. This may result in acceptance of QC results that do not meet acceptance criteria, which may cause erroneous patient results to be reported.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.