SIGNA Creator, SIGNA Explorer, NMRI systems

US FDA — Medical Devices · 2025-01-31 · ChinaClass II
Issuing authority
US FDA — Medical Devices
Alert date
2025-01-31
Country of alert
China
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
00195278577221
Brand on record
GE Healthcare (China) Co., Ltd.

What the authority reported

For certain MR systems, a scan can be resumed following a Power Monitor trip when the "6-minute average SAR" is above the limit indicated on the "SAR Display". The user interface may incorrectly enable a scan to proceed earlier than intended. If resumed by the operator, such a scan can result in potential excessive tissue heating, especially if the patient is in contact with a conductive material or has an implant.