Medtronic Pipeline Vantage Embolization Device with Shield Technology, REF: PED3-027-350-12, PED3-027-350-14, PED3-027-350-16, PED3-027-350-

US FDA — Medical Devices · 2025-01-30 · United StatesClass I
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2025-01-30
Country of alert
United States
Alert level
Class I
Alert type
recall
Barcode (EAN/GTIN)
00847536035263
Brand on record
Micro Therapeutics, Inc.

What the authority reported

Use of embolization device with diameters greater-than-or-equal-to 4mm may lead to incomplete wall apposition and/or braid deformation (fish-mouthing/braid narrowing/braid collapse) noted during procedure and post-procedure, typically noted at 6-12 month imaging follow-up, potentially can lead to thrombosis and/or serious adverse events. Higher risk in females less-than-or-equal-to 45 years of age