Official recall record
Brilliance iCT; Product Code (REF): 728306;
- Issuing authority
- US FDA — Medical Devices
- Recall initiated by firm
- 2026-01-07
- Country of alert
- United States
- Alert level
- Class II
- Alert type
- recall
- Barcode (EAN/GTIN)
- 00884838059474
- Brand on record
- Philips North America Llc
What the authority reported
As part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing within the gantry. One or more of these fasteners may not be torqued to specification. If multiple fasteners are not torqued to specification, the rotor or rotor parts may become unsecured or displaced. While there has been no reported or observed event of expelled parts during gantry rotation, unsecured rotor parts may potentially be expelled.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Record last checked against the registry on 2026-05-31.
Same barcode in other registries
The same EAN appears in 6 other official recall record(s) — a cross-border signal worth checking if you bought this product abroad.
- 2017-11-10 · US FDA — Medical Devices
Ingenuity Core 128, (Code 728323) Intended to produce images of the head and body by computer reconstruction of x-ray - 2020-04-06 · US FDA — Medical Devices
Philips Brilliance iCT Computed Tomography X-Ray System; Model # 728306; UDI #s (01)00884838059474(21)100783, (01)008848 - 2021-06-08 · US FDA — Medical Devices
728306 Brilliance iCT-Computed Tomography X-ray systems 728305 Brilliance iCT Upgrades - 2025-08-13 · US FDA — Medical Devices
Philips CT systems labeled as the following: 1. Brilliance iCT, Model Number: 728306. - 2017-11-10 · US FDA — Medical Devices
Brilliance iCT Intended to produce images of the head and body by computer reconstruction of x-ray transmission data
Other recalls from this brand
All 58 official recall notices naming Philips North America on this site →