HDS Traction Boot II, REF AR-6529-23, The Traction Boot is designed to position, support and/or distract the patient's foot for hip surgery
- Issuing authority
- US FDA — Medical Devices
- Alert date
- 2024-03-11
- Country of alert
- United States
- Alert level
- Class II
- Alert type
- recall
- Brand on record
- Baxter Healthcare Corporation
What the authority reported
Baxter Healthcare Corporation is issuing an Urgent Medical Device Recall for the Hip Distraction System, Traction Boot II, due to the boots being assembled incorrectly. The locking ridge on the front face of the boot key socket is facing upwards, which prevents the socket mechanism on the traction boot from attaching properly to the spar of the traction system. This will cause the boot to detach from the traction system.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Verify at the source:
US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.
Record last checked against the registry on 2026-05-31.
Other recalls from this brand
- 2026-08-18 · US FDA — Medical Devices
PRO+ 36" MRS SURFACE, mattress with Advanced Microclimate Technology (AMT) uses a non-powered system of intake - 2026-08-18 · US FDA — Medical Devices
PRO+ MRS SURFACE FOR VERSACARE, mattress with Advanced Microclimate Technology (AMT) uses a non-powered system - 2026-08-18 · US FDA — Medical Devices
PRO+ 40" MRS SURFACE, mattress with Advanced Microclimate Technology (AMT) uses a non-powered system of intake
ProductGuru Recalls republishes official safety information with attribution. We are not
the issuing authority and this page is not advice. If this product is a medical device or
medication, speak to a doctor or pharmacist. Found an error?
Report it — corrections are published.