ProductGuru Recalls

Official recall record

Situate Laparotomy Sponge, RF and X-ray Detectable, REF: L1818-04P01C-1

US FDA — Medical Devices · 2024-02-21 · United StatesClass II
Issuing authority
US FDA — Medical Devices
Alert date
2024-02-21
Country of alert
United States
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
20884521700953
Brand on record
Covidien
Product category
Medical Device

What the authority reported

Packs including 5 laparotomy sponges may contain incorrect product with a blue loop, X-ray detectable feature, and ethylene oxide treatment; however, this incorrect product wasn't steam pre-treated prior to sterilization, and has no RF tag, which if non-sterile, may lead to infection, tissue trauma, and sepsis, and if in the scannable area, would lead a "CLEAR" display.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: US FDA — Medical Devices · United States · all filters