VITROS Performance Verifier I

US FDA — Medical Devices · 2024-02-27 · United StatesClass II
Issuing authority
US FDA — Medical Devices
Alert date
2024-02-27
Country of alert
United States
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
10758750004317
Brand on record
Ortho-Clinical Diagnostics
Product category
Medical Device

What the authority reported

The issue was mismatched diluent lots assembled into VITRO Chemistry Products Performance Verifier I and II. This issue has the potential to cause erroneous Quality Control (QC) results that could lead to testing delays and erroneous patient results. This issue caused positive or negative bias in the quantification of sodium (Na+) and total carbon dioxide (ECO2). The incorrect diluent into the Performance Verifier II exhibited a potential negative bias of up to 6 mmol/L for ECO2 and Na+ testing.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: US FDA — Medical Devices · United States · all filters

Same barcode in other registries

The same EAN appears in 1 other official recall record(s) — a cross-border signal worth checking if you bought this product abroad.

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.