ProductGuru Recalls

Official recall record

Obsidio Conformable Embolic, Material Numbers (UPN): a) M0013972001010, b) M0013972101010

US FDA — Medical Devices · 2024-02-20 · United StatesClass I
Issuing authority
US FDA — Medical Devices
Alert date
2024-02-20
Country of alert
United States
Alert level
Class I
Alert type
recall
Barcode (EAN/GTIN)
00191506039332
Brand on record
Boston Scientific Corporation

What the authority reported

An investigation determined that delivery of the Obsidio embolic using the aliquot technique for lower gastrointestinal bleeding embolization poses a high risk of bowel ischemia. The most serious and the most common adverse health consequence, reasonably foreseeable to occur, is the need to perform major surgery such as bowel resection and/or diverting colostomy. Therefore, Boston Scientific does not recommend that the aliquot technique be used to deliver the Obsidio device for lower GI bleed embolization procedures.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: US FDA — Medical Devices · United States · all filters

Other recalls from this brand

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.