MONARCH Platform, Bronchoscopy, Catalog Numbers: MON-000005-01 , MON-000005-01R , MON-000006 , MON-000006-RFB, MON-000008 with software

US FDA — Medical Devices · 2026-01-21 · United StatesClass II
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2026-01-21
Country of alert
United States
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
10810068810803
Brand on record
Auris Health, Inc
Product category
Medical Device

What the authority reported

Device for bronchoscopic visualization, patient airway access has software issue: if application restarts after patient-side selection, prior to bronchoscope loading, application will re-initialize to patient-left position; and if right position was previously selected, this may lead to unanticipated robotic arm positioning toward patient-left, may result in contact with patient/endotracheal tube.