ProductGuru Recalls

Official recall record

Cardiovascular Procedure Kits that include High Crack Check Valve PN 205307A Part Number: 6375 16010 62974 65246 66312 66342 66388

US FDA — Medical Devices · 2024-02-20 · United StatesClass II
Issuing authority
US FDA — Medical Devices
Alert date
2024-02-20
Country of alert
United States
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
00699753001534
Brand on record
Terumo Cardiovascular Systems Corporation

What the authority reported

Pressure relief valve included in certain lots of cardiovascular procedure kits was reportedly opening at approximately 200-300mmHg versus 400mmHg, potential myocardial tissue damage potentially requiring surgical or medical/pharmacological intervention

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: US FDA — Medical Devices · United States · all filters