ARCHITECT Anti-HCV Reagent Kit, List Numbers: a) 1L79-25, b) 1L79-35

US FDA — Medical Devices · 2024-03-04 · GermanyClass II
Issuing authority
US FDA — Medical Devices
Alert date
2024-03-04
Country of alert
Germany
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
00380740001414
Brand on record
Abbott GmbH

What the authority reported

Customer complaints were received regarding falsely elevated results for some patient samples. Internal studies with customer return samples confirmed potential interactions that may lead to falsely elevated results when processed as below on the same instrument: " ARCHITECT Syphilis TP (LN 8D06) and/or ARCHITECT AdviseDx SARS-CoV-2-IgG II (LN 6S60) precedes ARCHITECT Anti-HCV testing (LN 1L79). " Alinity i Syphilis TP (LN 07P60) precedes Alinity i Anti-HCV testing (LN 08P05).