ProductGuru Recalls

Official recall record

EVO+ Visian Toric Implantable Collamer Lens (TICL), Model: VTICM5

US FDA — Medical Devices · 2026-01-21 · SwitzerlandClass II
Issuing authority
US FDA — Medical Devices
Alert date
2026-01-21
Country of alert
Switzerland
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
00840311304363
Brand on record
Staar Surgical AG

What the authority reported

Toric Implantable Collamer for correction/reduction of myopia in phakic eyes with astigmatism were labeled as 12.1 mm length but are 12.6 mm length lenses, so there is the potential for excessive vault, which may cause: Significant reduction of irido-corneal angles, Significant shallowing of anterior chamber, Patient discomfort, Elevated IOP/Angle closure, Unreactive pupil, Lens exchange/removal.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: US FDA — Medical Devices · Switzerland · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.