Official recall record
Owner's Booklets and Instructions for Use that are used with the following blood glucose measuring devices. Description/Item: Walgreens TR (ref Z-1547-2026)
- Issuing authority
- US FDA — Medical Devices
- Recall initiated by firm
- 2026-02-06
- Country of alert
- United States
- Alert level
- Class I
- Alert type
- recall
- Barcode (EAN/GTIN)
- 311917173894
- Brand on record
- Trividia Health, Inc.
What the authority reported
The system labeling (user manual and online labeling) did not provide adequate directions for lay users to act on the E-5 error code. The error code description and recommended actions for the E-5 error code in the labeling do not: (a) clearly emphasize that an E-5 error code could represent a very high blood glucose level, and (b) prominently convey the appropriate immediate actions, such as contacting a healthcare professional. This could lead to a delay in treatment if the user does not seek medical attention immediately when receiving an E-5 error code and experiencing symptoms of high glucose.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Record last checked against the registry on 2026-05-31.
Other recalls from this brand
- 2026-02-06 · US FDA — Medical Devices
Owner's Booklets and Instructions for Use that are used with the following blood glucose measurement devices. - 2026-02-06 · US FDA — Medical Devices
Owner's Booklets and Instructions for Use that are used with the following blood glucose measuring devices. D