ProductGuru Recalls

Official recall record

HeartMate Mobile Power Unit: REF: 107754, NA; 107758, EU; 107758AU, AU; 107758UK, UK; L107758, EU Rental. Used with HeartMate 3 and II Left

US FDA — Medical Devices · 2025-03-13 · United StatesClass I
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2025-03-13
Country of alert
United States
Alert level
Class I
Alert type
recall
Barcode (EAN/GTIN)
05415067038234
Brand on record
Thoratec LLC

What the authority reported

Left Ventricular Assist System Mobile Power Unit (MPU) may have electrical component that may cause MPU not to turn on, unprompted shut down, or turning off/restarting, may cause system controller to indicate Yellow Wrench or No External Power alarms. If MPU losses power, System Controller can support the pump up to 15 minutes, and if not switched to rechargeable batteries, pump may lose power.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.