Tandem pumps are battery-operated infusion pumps capable of both basal and bolus delivery of insulin. The pumps utilize a motor-driven mecha

US FDA — Medical Devices · 2026-01-30 · United StatesClass II
Issuing authority
US FDA — Medical Devices
Alert date
2026-01-30
Country of alert
United States
Alert level
Class II
Alert type
recall
Brand on record
Tandem Diabetes Care, Inc.

What the authority reported

The Czech language user guide contained multiple translation errors. The most significant error was that it incorrectly instructs users to verify that the infusion set is connected to the body, when it should instead direct users to confirm that the infusion set is not connected to the body. Following this instruction could result in unintended insulin delivery, potentially leading to a hypoglycemic event.