RefleXion X1 Radiotherapy System Model RXM1000, Part #800-00006-00
- Issuing authority
- US FDA — Medical Devices
- Recall initiated by firm
- 2026-02-09
- Country of alert
- United States
- Alert level
- Class II
- Alert type
- recall
- Barcode (EAN/GTIN)
- 00860003983812
- Brand on record
- Reflexion Medical, Inc.
What the authority reported
Due to a radiotherapy medical system software defect, when performing angular roll corrections followed by a repeat localization, the roll corrections applied in the initial localization scan are not carried through to treatment delivery leading to the treatment plan dose being delivered to the incorrect location, potentially a displacement of 5mm or higher, with a 10% to 20% underdose.