ProductGuru Recalls

Official recall record

Expo 5F Selective Angiographic Catheters, MODEL - 5F EXPO FL4.5 (5PK), REF H74908526232; cardiac catheter

US FDA — Medical Devices · 2024-03-21 · United StatesClass I
Issuing authority
US FDA — Medical Devices
Alert date
2024-03-21
Country of alert
United States
Alert level
Class I
Alert type
recall
Barcode (EAN/GTIN)
08714729187837
Brand on record
Boston Scientific Corporation

What the authority reported

An increase in complaints related to an inability to advance the guidewire through the lumen of the device for certain batches. There is evidence of the polyurethane layer delaminating and in some instances material detachment in the catheters inner lining.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: US FDA — Medical Devices · United States · all filters

Other recalls from this brand