Official recall record
Skytron Freedom 4FXS-60 Heavy Duty Monitor Mount
- Issuing authority
- US FDA — Medical Devices
- Recall initiated by firm
- 2025-03-27
- Country of alert
- Netherlands
- Alert level
- Class II
- Alert type
- recall
- Barcode (EAN/GTIN)
- 10841736116071
- Brand on record
- IHB OPERATIONS B.V.
- Product category
- Medical Device
What the authority reported
The mounting hardware that is supplied with the 4FXS-60 monitor bracket does not provide adequate thread engagement to support the weight of the frame and monitor. This could lead to a failure of the mounting hardware and subsequent separation of the frame from the mounting column.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Verify at the source:
US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.
Record last checked against the registry on 2026-05-31.