ProductGuru Recalls

Official recall record

Skytron Freedom 4FXS-60 Heavy Duty Monitor Mount

US FDA — Medical Devices · 2025-03-27 · NetherlandsClass II
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2025-03-27
Country of alert
Netherlands
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
10841736116071
Brand on record
IHB OPERATIONS B.V.
Product category
Medical Device

What the authority reported

The mounting hardware that is supplied with the 4FXS-60 monitor bracket does not provide adequate thread engagement to support the weight of the frame and monitor. This could lead to a failure of the mounting hardware and subsequent separation of the frame from the mounting column.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: US FDA — Medical Devices · Netherlands · all filters