Official recall record
Expo 5F Selective Angiographic Catheters, MODEL - 5F EXPO MULTIPACK 155 (5PKX3), REF H749085263022; cardiac catheter
- Issuing authority
- US FDA — Medical Devices
- Alert date
- 2024-03-21
- Country of alert
- United States
- Alert level
- Class I
- Alert type
- recall
- Barcode (EAN/GTIN)
- 08714729187882
- Brand on record
- Boston Scientific Corporation
What the authority reported
An increase in complaints related to an inability to advance the guidewire through the lumen of the device for certain batches. There is evidence of the polyurethane layer delaminating and in some instances material detachment in the catheters inner lining.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Verify at the source:
US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.
Record last checked against the registry on 2026-05-31.
Other recalls from this brand
- 2026-08-06 · US FDA — Medical Devices
LUX-Dx, LUX-Dx II, LUX-Dx II+ Insertable Cardiac Monitor System, SERVER SW LAT CLARITY DS ANZ, Model Number 72 - 2026-08-06 · US FDA — Medical Devices
LUX-Dx, LUX-Dx II, LUX-Dx II+ Insertable Cardiac Monitor System, SERVER SW LAT CLARITY DS US, Model Number 726 - 2026-08-06 · US FDA — Medical Devices
LUX-Dx, LUX-Dx II, LUX-Dx II+ Insertable Cardiac Monitor System, SERVER SW LAT CLARITY DS EU, Model Number 724
ProductGuru Recalls republishes official safety information with attribution. We are not
the issuing authority and this page is not advice. If this product is a medical device or
medication, speak to a doctor or pharmacist. Found an error?
Report it — corrections are published.