Official recall record
Medline kits containing Olympus biopsy valves: 1. BRONCH PROCEDURE KIT, DYKE2091 2. BRONCHSCOPY, DYNJ900898G DYNJ901922I 3. BRONC
- Issuing authority
- US FDA — Medical Devices
- Recall initiated by firm
- 2026-02-16
- Country of alert
- United States
- Alert level
- Class II
- Alert type
- recall
- Barcode (EAN/GTIN)
- 10198459606397
- Brand on record
- Medline Industries, LP
What the authority reported
Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Record last checked against the registry on 2026-05-31.
Other recalls from this brand
- 2026-08-14 · US FDA — Medical Devices
Rollator, Bariatric, Basic Steel, Blue, mechanical walker with four legs and a metal frame intended for medica - 2026-08-14 · US FDA — Medical Devices
Rollator, Bariatric, Basic Steel, Burgundy, mechanical walker with four legs and a metal frame intended for me - 2026-08-14 · US FDA — Medical Devices
Medline Basic Homecare Beds and Medlite Homecare Beds and Accessories, Hand Pendants: MDR510094 PENDANT F/
All 257 official recall notices naming Medline Industries, LP on this site →