Official recall record
PiezoWave 2 Control Unit, Product Number 100506US, compact myofascial acoustic compression therapy device that enables treatment with piezoe
- Issuing authority
- US FDA — Medical Devices
- Recall initiated by firm
- 2025-04-09
- Country of alert
- Germany
- Alert level
- Class II
- Alert type
- recall
- Barcode (EAN/GTIN)
- 00814334022863
- Brand on record
- Richard Wolf GmbH
What the authority reported
The high voltage capacitor may fail early in the product life cycle after delivery of surge voltages. When this failure occurs, the capacitor may produce smoke and a burning smell. In extremely rare instances a capacitor may cause a spark or excessive heat when an internal capacitor failure occurs.