Official recall record
SmartPath to dStream for 1.5T, Model Number: 782146;
- Issuing authority
- US FDA — Medical Devices
- Recall initiated by firm
- 2025-03-24
- Country of alert
- Netherlands
- Alert level
- Class II
- Alert type
- recall
- Brand on record
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
What the authority reported
The Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Verify at the source:
US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.
Record last checked against the registry on 2026-05-31.
Other recalls from this brand
- 2026-07-16 · US FDA — Medical Devices
Philips Azurion Rystem, Model Numbers: 722063, 722064, 722067, 722068, 722078, 722079, 722221, 722222,722223 - 2026-07-16 · US FDA — Medical Devices
Philips Azurion, Model Numbers: 722229, 722230, 722231, 722232, 722233, 722234, 722235, and 722236, with Soft
All 73 official recall notices naming PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. on this site →
ProductGuru Recalls republishes official safety information with attribution. We are not
the issuing authority and this page is not advice. If this product is a medical device or
medication, speak to a doctor or pharmacist. Found an error?
Report it — corrections are published.