Intelli-C: Right side suspended Tabletop, REF: 03400000; Left side suspended Tabletop, REF: 03400010, used with the Intelli-C X-ray system
- Issuing authority
- US FDA — Medical Devices
- Alert date
- 2024-04-17
- Country of alert
- Denmark
- Alert level
- Class II
- Alert type
- recall
- Brand on record
- NRT X-RAY A/S
What the authority reported
X-ray system C-arm may experience uncontrolled motion if 1) AC motor controller firmware has a faulty setting, 2) the Motor/gear assembly is worn, 3) CAT movement is driven at maximum speed and angulated more than 75 deg. from vertical, and 4) the emergency stop or touch guard is activated during the breaking sequence, which could lead to the C-arm hitting the patient or operator.