Intelli-C: Right side suspended Tabletop, REF: 03400000; Left side suspended Tabletop, REF: 03400010, used with the Intelli-C X-ray system

US FDA — Medical Devices · 2024-04-17 · DenmarkClass II
Issuing authority
US FDA — Medical Devices
Alert date
2024-04-17
Country of alert
Denmark
Alert level
Class II
Alert type
recall
Brand on record
NRT X-RAY A/S

What the authority reported

X-ray system C-arm may experience uncontrolled motion if 1) AC motor controller firmware has a faulty setting, 2) the Motor/gear assembly is worn, 3) CAT movement is driven at maximum speed and angulated more than 75 deg. from vertical, and 4) the emergency stop or touch guard is activated during the breaking sequence, which could lead to the C-arm hitting the patient or operator.