Evidence MultiSTAT with software v 3.7-Analyzer intended for the qualitative determination of parent drug molecule and metabolites of drugs

US FDA — Medical Devices · 2024-04-05 · IrelandClass III
Issuing authority
US FDA — Medical Devices
Alert date
2024-04-05
Country of alert
Ireland
Alert level
Class III
Alert type
recall
Barcode (EAN/GTIN)
05055273214970
Brand on record
Randox Laboratories Ltd.

What the authority reported

Foam gasket on the Evidence MultiSTAT chip heater assembly may not perform as expected. Foam that is not functioning as expected may cause light to leak into the Charged Coupled Device (CCD) Camera during imaging. This will affect the testing and may delay in reporting results.