Apo B Reagent, REF: OSR6143

US FDA — Medical Devices · 2025-04-23 · IrelandClass III
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2025-04-23
Country of alert
Ireland
Alert level
Class III
Alert type
recall
Barcode (EAN/GTIN)
15099590010409
Brand on record
Beckman Coulter Ireland, Inc.
Product category
Medical Device

What the authority reported

The product Value Assignment Sheet (VAS) is used in calibration of the device. An erroneous value was included in the VAS which has the potential to generate an incorrect APO B patient result which could exceed the total allowable error for the APO B assay (11.6%). The error results in results with a positive bias up to 42%. However, based on the mitigating factors the risk to health is considered to be negligible or limited.