Airvo 2 Humidifier, REF: PT101US, and myAirvo 2 Humidifier, REF: PT100US

US FDA — Medical Devices · 2024-04-02 · New ZealandClass II
Issuing authority
US FDA — Medical Devices
Alert date
2024-04-02
Country of alert
New Zealand
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
09420012422248
Brand on record
Fisher & Paykel Healthcare, Ltd.

What the authority reported

Humidifier devices, used to deliver high flow respiratory therapy to patients, manufactured before 14 August 2017, have a speaker configuration that may result in distorted, intermittent or inaudible alarm sound levels. In the absence of an audible alarm, if there is an interruption to therapy, a patient may experience oxygen desaturation.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.