ProductGuru Recalls

Official recall record

The Sophysa Pressio Ventricular Intracranial Pressure Monitoring Kit monitoring is intended for monitoring intracranial ventricular pressure (ref Z-1826-2024)

US FDA — Medical Devices · 2024-04-09 · FranceClass II
Issuing authority
US FDA — Medical Devices
Alert date
2024-04-09
Country of alert
France
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
3760124132922
Brand on record
Sophysa S.A.
Product category
Medical Device

What the authority reported

An increase in claims related to leakage of cerebrospinal fluid (CSF) from the proximal end of the catheter. In the event of a leak going undetected during implantation, it could pose an infectious risk to the patient. The cause of the leaks is indicated to be manufacturing error involving perforation of the inner channel of the catheter during stylet insertion.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: US FDA — Medical Devices · France · all filters