ProductGuru Recalls

Official recall record

Broselow Pediatric Emergency Rainbow Tape (distribution by AirLife); REF 7700REA AirLife brand, 2025 Edition, and 36-23446 Rev 2 Print V

US FDA — Medical Devices · 2025-05-15 · United StatesClass I
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2025-05-15
Country of alert
United States
Alert level
Class I
Alert type
recall
Barcode (EAN/GTIN)
10889483588970
Brand on record
SunMed Holdings, LLC
Product category
Medical Device

What the authority reported

The impacted tape was manufactured with incorrect information on the tape. Incorrect values are printed in the Red zone, Orange zone, and Grey zone. Using the incorrect values in the Red zone could lead to shocking a patient with an excessive dose of joules, causing significant harm including burns, heart damage, and cardiac arrest. Incorrect values in the Orange zone could lead to overdosing the patient with sodium bicarbonate and cause metabolic alkalosis, electrolyte imbalances, tissue damage, and worsen respiratory status. Incorrect information in the Grey may lead to underdosing the patient with sodium bicarbonate and may cause reduced myocardial contractility, decreased response to vasopressors, and increased risk of dysrhythmia.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: US FDA — Medical Devices · United States · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.