ProductGuru Recalls

Official recall record

MIRA SURGICAL SYSTEM, Table mounted miniaturized electromechanical surgical system, Catalog Number MIRA-CAP-110, Software Version 4.0.3

US FDA — Medical Devices · 2025-04-23 · United StatesClass II
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2025-04-23
Country of alert
United States
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
00850038042028
Brand on record
VIRTUAL INCISION CORPORATION

What the authority reported

The MIRA Surgical System Software Version 4.0.3 loaded on the Surgeon Control Console may have the potential to exhibit interrupted motion on the display screen, caused by a software latency between the movement of the hand controllers and the movement of the Surgical Minibot arms.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.