ProductGuru Recalls

Official recall record

DAC 044 115cm, Catalog: 90760

US FDA — Medical Devices · 2024-04-23 · United StatesClass II
Issuing authority
US FDA — Medical Devices
Alert date
2024-04-23
Country of alert
United States
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
07613327313918
Brand on record
Stryker Neurovascular

What the authority reported

Neurovascular devices did not meet the minimum sample size required for endotoxin testing during manufacturing processes and are out of compliance with the AAMI ST72 standard, so these devices may exceed the acceptable limits of endotoxin levels; high levels of endotoxins may increase the risk of fever, shock and aseptic meningitis.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.