Official recall record
Olympus EVIS EXERA III Duodenovideoscope. Model Number: TJF-Q190V
- Issuing authority
- US FDA — Medical Devices
- Recall initiated by firm
- 2025-05-05
- Country of alert
- United States
- Alert level
- Class II
- Alert type
- recall
- Barcode (EAN/GTIN)
- 04953170405563
- Brand on record
- Olympus Corporation of the Americas
What the authority reported
Recent post-market surveillance data suggests a possible association of higher microbial contamination levels from TJF duodenoscopes when manual cleaning was delayed beyond one hour and a presoak was performed, compared to those TJF duodenoscopes where manual cleaning began within one hour after patient procedure.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Record last checked against the registry on 2026-05-31.
Same barcode in other registries
The same EAN appears in 1 other official recall record(s) — a cross-border signal worth checking if you bought this product abroad.
- 2023-07-27 · US FDA — Medical Devices
EVIS EXERA III DUODENOVIDEOSCOPE OLYMPUS TJF-Q190V
Other recalls from this brand
All 8 official recall notices naming Olympus Corporation of the Americas on this site →