Official recall record
Intera 1.5T Explorer/Nova Dual Product Number: 781108;
- Issuing authority
- US FDA — Medical Devices
- Recall initiated by firm
- 2025-05-29
- Country of alert
- United States
- Alert level
- Class II
- Alert type
- recall
- Brand on record
- Philips North America
What the authority reported
identified an issue where a component failure in the specific Gradient Coil type of affected systems can potentially act as a heat source, with a potential to produce smoke and/or fire. If smoke or a fire occurs, the risk to patients or operators may include inhalation of smoke, burns, and/or asphyxia which may lead to injury or even death. This issue could also lead to property damage.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Verify at the source:
US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.
Record last checked against the registry on 2026-05-31.
Other recalls from this brand
- 2026-08-28 · US FDA — Medical Devices
Ingenia 1.5T, 1.5T CX, and 1.5T S; Model number (REF): 1. 781315 (1.5T), 2. 781341 (1.5T), 3. 781396 (1.5T), - 2026-08-28 · US FDA — Medical Devices
MR 7700; Model number (REF): 1. 782120, 2. 782153, 3. 782155, 4. 782199; Software Versions: R11, R12 and R13 - 2026-08-28 · US FDA — Medical Devices
Evolution upgrade 1.5T amd Evolution upgrade 3.0T; Model number (REF): 1. 782116 (1.5T), 2. 782148 (1.5T), 3.
All 58 official recall notices naming Philips North America on this site →
ProductGuru Recalls republishes official safety information with attribution. We are not
the issuing authority and this page is not advice. If this product is a medical device or
medication, speak to a doctor or pharmacist. Found an error?
Report it — corrections are published.