VOCSN V+Pro package, REF: PRT-01198-000; VOCSN V+Pro unit, REF: PRT-01185-000

US FDA — Medical Devices · 2026-03-25 · United StatesClass I
Issuing authority
US FDA — Medical Devices
Alert date
2026-03-25
Country of alert
United States
Alert level
Class I
Alert type
recall
Barcode (EAN/GTIN)
00855573007914
Brand on record
Ventec Life Systems, Inc.

What the authority reported

Respiratory system intended to provide continuous/intermittent ventilator support, may not have been evaluated under all intended high-pressure conditions during production testing, so an oxygen leak condition could occur that may lead to reduced below intended levels of fraction of inspired oxygen delivery prior to and during ventilation and this oxygen-enriched environment may increase fire risk