ProductGuru Recalls

Official recall record

Dexcom One+ Continuous Glucose Monitoring System, SKU(Part Number): STK-D7-013 (MT28285-1), STK-D7-014 (MT28285-1), STK-D7-103 (MT28285-2)

US FDA — Medical Devices · 2025-05-12 · United StatesClass I
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2025-05-12
Country of alert
United States
Alert level
Class I
Alert type
recall
Barcode (EAN/GTIN)
00386270003812
Brand on record
Dexcom, Inc.

What the authority reported

Defective foam or an assembly error may cause the receiver speaker to lose contact with the printed circuit board, leading to missed audible alerts for low or high blood glucose values. A missed audible alert for low or high blood glucose values could lead to untreated hypo or hyperglycemia which can cause seizures, vomiting, loss of consciousness, or death.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: US FDA — Medical Devices · United States · all filters

Other recalls from this brand

All 13 official recall notices naming Dexcom, Inc. on this site →

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.