Official recall record
Newport s N-LHP-928(j) units if sold as a laser head/power supply set, and N5-LPL-944-080 and N6-LPL-944-080 if sold as a power supply alone
- Issuing authority
- US FDA — Medical Devices
- Alert date
- 2024-03-27
- Country of alert
- United States
- Alert level
- Class II
- Alert type
- recall
- Brand on record
- Newport Corp
What the authority reported
According to 21 CFR 1040.10(f)(4), each laser system classified as a Class IIIb or IV laser product shall incorporate a key-actuated master control. The key shall be removable and the laser shall not be operable when the key is removed. However, the key of the subject laser products can be removed from key switch and the laser remains on. This is a safety concern.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Verify at the source:
US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.
Record last checked against the registry on 2026-05-31.
ProductGuru Recalls republishes official safety information with attribution. We are not
the issuing authority and this page is not advice. If this product is a medical device or
medication, speak to a doctor or pharmacist. Found an error?
Report it — corrections are published.