Official recall record
Philips Ingenia 3.0T with MR Elastography (MRE). 1. Model Number (REF): 781342. 2. Model Number (REF): 781377. 3. Model Number (
- Issuing authority
- US FDA — Medical Devices
- Alert date
- 2026-04-14
- Country of alert
- United States
- Alert level
- Class II
- Alert type
- recall
- Barcode (EAN/GTIN)
- 00884838055339
- Brand on record
- Philips North America
What the authority reported
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Verify at the source:
US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.
Record last checked against the registry on 2026-05-31.
Same barcode in other registries
The same EAN appears in 4 other official recall record(s) — a cross-border signal worth checking if you bought this product abroad.
- 2025-06-30 · US FDA — Medical Devices
Ingenia 3.0T Model Numbers (REF): (1) 781342, (2) 781377, (3) 782103; - 2025-12-03 · US FDA — Medical Devices
Ingenia 3.0T. Product Code (REF): (1) 781342, (2) 781377. MR systems with SW version R11.1 and R12.1. - 2025-05-13 · US FDA — Medical Devices
Product Name: Ingenia 3.0T; Model Numbers: (1) 781342, (2) 781377, (3) 782103; - 2021-07-30 · US FDA — Medical Devices
Ingenia 3.0T-Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can pro
Other recalls from this brand
- 2026-04-14 · US FDA — Medical Devices
Philips Achieva 1.5T with MR Elastography (MRE). 1. Model Number (REF): 781196. 2. Model Number (REF): - 2026-04-14 · US FDA — Medical Devices
Philips Ingenia 1.5T CX with MR Elastography (MRE). 1. Model Number (REF): 781262. - 2026-04-14 · US FDA — Medical Devices
Philips Ingenia Ambition S with MR Elastography (MRE). 1. Model Number (REF): 781359. 2. Model Number
ProductGuru Recalls republishes official safety information with attribution. We are not
the issuing authority and this page is not advice. If this product is a medical device or
medication, speak to a doctor or pharmacist. Found an error?
Report it — corrections are published.