ProductGuru Recalls

Official recall record

Baxter Novum IQ Syringe Pump, product code 40800BAXUS,

US FDA — Medical Devices · 2025-05-12 · United StatesClass II
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2025-05-12
Country of alert
United States
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
05413765852428
Brand on record
Baxter Healthcare Corporation

What the authority reported

Baxter has identified a design issue which may cause the pump to detect the presence of a syringe flange when no syringe is loaded. When this issue occurs, the pumps 'user interface becomes stuck on the "Remove Syringe" screen when an infusion is complete, and the syringe is unloaded. A System Error 21502 (Flange Sensor Failure) may occur. Please note this issue only affects a subset of Novum IQ syringe pumps manufactured or repaired after July 4, 2023.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: US FDA — Medical Devices · United States · all filters

Same barcode in other registries

The same EAN appears in 2 other official recall record(s) — a cross-border signal worth checking if you bought this product abroad.

Other recalls from this brand

All 43 official recall notices naming Baxter Healthcare Corporation on this site →