Official recall record
Heraeus, PALAMIX duo. Material Number: 66057897.
- Issuing authority
- US FDA — Medical Devices
- Alert date
- 2026-03-25
- Country of alert
- Germany
- Alert level
- Class III
- Alert type
- recall
- Brand on record
- Heraeus Medical GmbH (Dental Division)
What the authority reported
Based on long-term stability testing, it was determined that in units older than 1.5 years (17 months) the mixing rod may, in rare cases, come loose from its lock during the mixing process. Notification is to reduce the shelf life from 3 years to 1.5 years (17 months).
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Verify at the source:
US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.
Record last checked against the registry on 2026-05-31.
Other recalls from this brand
- 2026-03-25 · US FDA — Medical Devices
Heraeus, PALAMIX uno. Material Number: 66057893.
ProductGuru Recalls republishes official safety information with attribution. We are not
the issuing authority and this page is not advice. If this product is a medical device or
medication, speak to a doctor or pharmacist. Found an error?
Report it — corrections are published.