Official recall record
Product Name [REF]: Supply Server, CCE Enterprise SW Site License (2-5), BD Pyxis Logistics [806], BD Pyxis MedStation 4000 Main 10885403512
- Issuing authority
- US FDA — Medical Devices
- Recall initiated by firm
- 2025-05-08
- Country of alert
- United States
- Alert level
- Class II
- Alert type
- recall
- Barcode (EAN/GTIN)
- 10885403512629
- Brand on record
- CareFusion 303, Inc.
- Product category
- Medical Device
What the authority reported
During automated dispensing cabinet upgrade/installation/reimaging, component manager was left in installed mode, and if a Microsoft patch is then installed this may lead to an error preventing user access to the dispensing software application, which may lead to delay in access to medications/supplies, and delivery/replenishment of patient specific orders; which may lead to delay in patient care.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Record last checked against the registry on 2026-05-31.
Same barcode in other registries
The same EAN appears in 4 other official recall record(s) — a cross-border signal worth checking if you bought this product abroad.
- 2024-04-15 · US FDA — Medical Devices
BD Pyxis MedStation 4000 System, REF: 303, Medication Cabinet - 2023-02-14 · US FDA — Medical Devices
BD Pyxis MedStation 4000 Main, REF 303 - 2024-02-23 · US FDA — Medical Devices
BD Pyxis MedStation 4000 REF: 303. Intended to securely store and dispense medications to a qualified and authorized he - 2026-04-30 · US FDA — Medical Devices
BD Pyxis MedStation 4000 REF: 303 Medication cabinet
Other recalls from this brand
All 11 official recall notices naming CareFusion 303, Inc. on this site →