Official recall record
MRI systems: Vantage Orian MRT-1550 (MEXL-1550), Vantage Elan MRT-2020 (MEXL-1520), Vantage Titan MRT-1504 (MEXL-1504), Vantage Titan 3T MRT
- Issuing authority
- US FDA — Medical Devices
- Alert date
- 2024-05-10
- Country of alert
- United States
- Alert level
- Class II
- Alert type
- recall
- Barcode (EAN/GTIN)
- 04987670103360
- Brand on record
- Canon Medical System, USA, INC.
What the authority reported
The terminal block used to secure the power supply cable for the gradient coil of the MRI system was not secured as designed, which could cause cable disconnection, which could lead to arcing, followed by melting of the cable, ignition, smoke, and fire.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Verify at the source:
US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.
Record last checked against the registry on 2026-05-31.
Other recalls from this brand
- 2025-12-19 · US FDA — Medical Devices
Alphenix INFX-8000F, interventional fluoroscopic x-ray system - 2025-12-19 · US FDA — Medical Devices
VANTAGE GALAN 3T Model MRT-3020/MEXL-3020 - 2025-12-19 · US FDA — Medical Devices
VANTAGE TITAN 3T Model MRT-3010/MEXL-3010
ProductGuru Recalls republishes official safety information with attribution. We are not
the issuing authority and this page is not advice. If this product is a medical device or
medication, speak to a doctor or pharmacist. Found an error?
Report it — corrections are published.