Official recall record
The Spectrum IQ Infusion System with Dose IQ Safety Software, Product Code 3570009
- Issuing authority
- US FDA — Medical Devices
- Recall initiated by firm
- 2023-06-15
- Country of alert
- United States
- Alert level
- Class I
- Alert type
- recall
- Barcode (EAN/GTIN)
- 00085412610900
- Brand on record
- Baxter Healthcare Corporation
What the authority reported
A medical device correction is being issued for SIGMA Spectrum Infusion System (V8 Platform) and Spectrum IQ Infusion System with Dose IQ Safety Software infusion pumps due to an increase in reported false upstream occlusion alarms following upgrades to software versions v8.01.01 and v9.02.01.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Record last checked against the registry on 2026-05-31.
Same barcode in other registries
The same EAN appears in 6 other official recall record(s) — a cross-border signal worth checking if you bought this product abroad.
- 2025-07-14 · US FDA — Medical Devices
Spectrum IQ Infusion Pump, Product Code 3570009 - 2024-10-02 · US FDA — Medical Devices
Baxter Spectrum IQ Infusion System with Dose IQ Safety Software, Product Code 3570009 - 2018-09-18 · US FDA — Medical Devices
SPECTRUM IQ INFUSION SYSTEM WITH DOSE IQ SAFETY SOFTWARE, 3570009 - 2021-12-29 · US FDA — Medical Devices
Baxter Spectrum IQ Infusion Pumps, Product code 3570009. - 2026-01-13 · US FDA — Medical Devices
Brand Name: Spectrum IQ Product Name: Infusion Pump Model/Catalog Number: 3570009 Software Version: N/A Product Desc
Other recalls from this brand
- 2026-09-14 · US FDA — Medical Devices
Likorall Overhead Lifts Product Code: 3121001 Product Description: LIKORALL 200 Product Code: 3122005 - 2026-08-18 · US FDA — Medical Devices
PRO+ 36" MRS SURFACE, mattress with Advanced Microclimate Technology (AMT) uses a non-powered system of intake - 2026-08-18 · US FDA — Medical Devices
PRO+ MRS SURFACE FOR VERSACARE, mattress with Advanced Microclimate Technology (AMT) uses a non-powered system
All 48 official recall notices naming Baxter Healthcare Corporation on this site →