ProductGuru Recalls

Official recall record

Impella LD intravascular micro axial blood pump, Product Number 005082

US FDA — Medical Devices · 2023-06-14 · United StatesClass I
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2023-06-14
Country of alert
United States
Alert level
Class I
Alert type
recall
Barcode (EAN/GTIN)
00813502011227
Brand on record
Abiomed, Inc.

What the authority reported

There is a potential risk for unintentional interaction of the Impella motor housing with the distal stent of a transcatheter aortic valve replacement (TAVR) resulting in destruction of the impeller blades. This has resulted in low flow from the damaged Impella system. Systemic embolization of the fractured impeller material is a possibility.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: US FDA — Medical Devices · United States · all filters

Other recalls from this brand

All 38 official recall notices naming Abiomed, Inc. on this site →