Cardiosave Rescue Intra-Aortic Balloon Pump (IABP), Model numbers 0998-00-0800-75 0998-00-0800-83 0998-00-0800-85

US FDA — Medical Devices · 2023-06-05 · United StatesClass I
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2023-06-05
Country of alert
United States
Alert level
Class I
Alert type
recall
Barcode (EAN/GTIN)
10607567112312
Brand on record
Datascope Corp.
Product category
Medical Device

What the authority reported

IABP may lose the ability to charge batteries in one or both bay slots. Therapy may be interrupted if batteries fail to charge and the device is disconnected from AC power. Low battery alarms may alert the User to the issue prior to interruption of therapy.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: US FDA — Medical Devices · United States · all filters

Same barcode in other registries

The same EAN appears in 6 other official recall record(s) — a cross-border signal worth checking if you bought this product abroad.

Other recalls from this brand

All 14 official recall notices naming Datascope Corp. on this site →

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.